16
Powering Your PAP Device: ResMed S8 Devices
ResMed S8 Series PAP device input ports run opposite polarity from other
manufacturers so the use of ResMed’s S8 Series DC-12 Converter or a
sine wave power inverter is required to complete the connection between
the device and the Battery Pack. DO NOT connect the Battery Pack
directly to any ResMed model device.
Insert the DC input cord into the DC out jack of the Battery Pack.
DUAL BATTERY PACK CONFIGURATION:
Insert the DC power pigtail cord into the DC out jacks of
both Battery Packs.
Insert the DC power pigtail cord into the DC input cord.
Insert the ResMed DC-12 Converter into the DC input cord and then
into the DC input port of the S8 PAP device.
Turn the PAP device on and it should begin functioning.
PAP DC INPUT PORT
DC OUT JACK
DC INPUT CORD
RESMED S8 SERIES
DC-12 CONVERTER
Note: A sine wave power inverter may also be used to complete the connection between
the ResMed S8 device and the Battery Pack.